Johnson
Noise Thermometry Results
The noise at the output of Metrosol’s first high
temperature JNT system was assessed whilst immersed in a
block bath at ambient, 800 & 1,000°C. At 1,000°C
it was found to be ~1.1 °C.
The theoretical limit for JNT (given by Rice's equation)
is also shown for the bandwidth of the electronics (1 MHz)
and sampling interval (5.872s). Note: the effective JNT
bandwidth will be less than 1 MHz due to RC filtering in
the probe. The JNT result could be improved by weighting
local frequency contributions by their signal to noise ratio.
The standard deviation from a thermocouple is also shown
at 800 & 1,000 °C.


High temperature
standard deviations for JNT thermocouple and Rices
equation 2 |
|